AMS MACHINE



AMS Laboratory in Poznań is equipped with the 1.5 SDH-Pelletron Model "Compact Carbon AMS" ser. no. 003, produced in 2001 by the National Electrostatics Corporation , Middleton, USA, and purchased by the Adam Mickiewicz University in Poznań (Poland). 

akcelcz1a.png akcelcz2a.png akcelcz3a.png
akcelcz4a.png
The parts of the AMS spectrometer are placed on sides of a square 5x5m large. Assembled, they form one high-vacuum chamber, about 15m long.

akcelcz6a.png
akcelcz7a_copy.png akcelcz8a.png akcelcz9b.png



To determine carbon isotopic ratios (14C/12C and 14C/13C), the AMS machine simultaneously measures number of 12C, 13C and 14C atoms released from cathode earlier produced from the sample .
The AMS machine produces carbon ions in the ion source, and injects them into accelerator. To avoid interference with other elements, ions of desired mass are separated in the injection magnet. The injected ions (12C-, 13C- and 14C-) are pulled into the accelerator terminal (charged to potential of +500,000V), there transform into 12C+, 13C+ and 14C+, and are pushed away, towards analysing magnet. This magnet directs 12C+ and 13C+ ions into specified spots in the drift chamber, which measure electric current of both types of ions. The rare ions of 14C+, after passing electrostatic analyser, hit the detector sensitive to count them.
All essential parts and parameters of the spectrometer, during maintenance and measurements are computer controlled and may be set from the console.
 
le_he.png
Low-energy side of the AMS spectrometer
High-energy side of the AMS spectrometer


 